This snapshot is older than v21 and keeps notes and UI only; the large CAD, mesh and render files are no longer on this site.
New in this snapshot: 239 drawings, 0 STEP files, 0 documents/media.
2026-10-28 9:30 PM EDT — Run 14: 3 in tube + Luer-lock tip baseline
- Split the 4 in reach into a 304 SS 4×0.5 tube (76.2 long) and a male Luer-lock adapter (ISO 80369-7) for standard 0.25 in Luer-lock dispensing tips. A 25 ga tip deflects 37 µm at 0.1 N; f1 is 277 Hz.
- Added 2 bronze sleeves (4×7×8) as a tube guide, because slider play would otherwise give 1.55 mm at the tip. The ≤ 0.05 mm runout target needs up-camera offset calibration; mechanically it is about 0.1–0.2 mm.
- Head clash check over 14 poses: 0 hits. Integration sweep over 206 poses: HOME at board Z has 0 hits for T 322–540. The N4 needle-down rule T ≥ 462 still holds.
- Exported STEP (home, N1–N6, components), GLB, USDZ and the nozzle close-up render. Plan section 10.7, CAM_HEAD_V3.md and the BOM are updated. Not deployed.
2026-10-29 4:21 PM EDT — Approvals (the designer)
- APPROVED: the hot-air (E) nozzle stays on the head-front bracket.
- APPROVED: E-only Z soft limit Z_MODE_E = 327.7, allowed only at T >= 462. Other stations keep Z_MIN_SOFT 354.7.